MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/10/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number: 1627487-2020-47693, 1627487-2020-47695, 1627487-2020-47696.It was reported during same surgery that occurred (b)(6) 2019, wherein a full drg system was explanted, that a portion of one lead that broke was confirmed on (b)(6) 2020 by x-ray remains implanted.Patient remains stable.The drg system explant is documented in related manufacturer reference numbers 1627487-2019-13872, 1627487-2019-13078,1627487-2019-13880, 1627487-2019-13881.
 
Manufacturer Narrative
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
Additional information received indicated that no further intervention is planned for this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key10909018
MDR Text Key218509159
Report Number1627487-2020-47694
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/17/2020
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number6265443
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2020
Initial Date FDA Received11/26/2020
Supplement Dates Manufacturer Received12/29/2020
Supplement Dates FDA Received01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DRG SLIM TIP LEAD (X3)
Patient Outcome(s) Other;
Patient Weight104
-
-