• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII NON POR FIN PUNCH 7-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII NON POR FIN PUNCH 7-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440270
Device Problems Dull, Blunt (2407); Sharp Edges (4013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Event Description
It was reported that, during set up, the device was dull.The procedure finished with the same device.No surgical delay.Patient injuries were not reported.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The cutting edge of the device is dull and worn rendering it inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII NON POR FIN PUNCH 7-8
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10909083
MDR Text Key218638687
Report Number1020279-2020-06734
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010214775
UDI-Public03596010214775
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/02/2006
Device Model Number71440270
Device Catalogue Number71440270
Device Lot Number61000534
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2020
Date Manufacturer Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-