The initial reporter received discrepant results for one patient sample tested for potassium on the roche 9180 electrolyte analyzer.The patient's potassium results were not reported outside the laboratory.The customer performed repeat testing with the patient's sample for confirmation.At 12:40 p.M., the patient's potassium result was 5.4 mmol/l with a data flag.At 12:42 p.M., the patient's potassium result was 5.6 mmol/l with a data flag.At 12:59 p.M., the patient's potassium result was 5.3 mmol/l with a data flag.At 13:01 p.M., the patient's potassium result was 4.5 mmol/l.At 13:02 p.M., the patient's potassium result was 5.3 mmol/l with a data flag.The customer performed an additional potassium measurement, and the patient's potassium result was 5.1 with a data flag.The time of measurement was requested but not provided.The potassium electrode lot number was 21500348 with an expiration date requested but not provided.The potassium electrode had an install by date of(b)(6) 2020, and the electrode was installed on (b)(6) 2020.
|
The potassium electrode lot number was 21500348 with an expiration date of 31-jul-2020.The investigation confirmed the customer performed the recommended maintenance procedures on the system.The investigation requested the affected snappak, reference electrode, potassium electrode, and needle unit for further investigation.The customer was unable to send the parts in for an investigation as the parts remained in use.The customer has not reported any further issues.The investigation did not identify a product problem.The cause of the event could not be determined.
|