• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C-US
Device Problems Break (1069); Positioning Failure (1158); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2020
Event Type  malfunction  
Event Description
It was reported that the distal filter inadvertently deployed during device removal.A sentinel cerebral protection system was used during a transcatheter aortic valve replacement (tavr) procedure.After deployment of the sentinel cerebral protection proximal filter, the distal filter could not be deployed.The procedure was continued with only the proximal filter deployed.At some point during the procedure, the #3 distal filter hypotube broke off of the handle of the sentinel cerebral protection system.When the physician was removing the sentinel device from the patient, the distal filter was inadvertently deployed in the patients radial artery.The distal filter was recaptured and the sentinel cerebral protection system was removed from the patient together with the introducer sheath.No patient harm occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
CLARET MEDICAL, INC.
1745 copperhill parkway
santa rosa CA 95403
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
160 knowles drive
los gatos CA 95032
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10924519
MDR Text Key218934823
Report Number2134265-2020-16485
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/10/2022
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0025997745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-