• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 195215
Device Problems Connection Problem (2900); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2020
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) could not verify the problem.The yellow alert sensor was slightly worn, but the unit tested normally.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, a level sensor 'detached' message displayed on the central control monitor (ccm).No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
Per the field service representative, the user facility's manufactured trained biomedical technician replaced the worn level sensor.
 
Manufacturer Narrative
Per the field service representative, the user facility's manufacturer trained biomedical technician ordered a new level sensor and replaced the old level sensor.
 
Manufacturer Narrative
Updated block: b5 per data log analysis, the level log shows many alert probe status changes from coupled to uncoupled.This will cause a 'level not attached' message and a 'level not attached alert' if the level detection is turned on.This usually indicates a problem with the level pad placement or the probe itself (less likely).The log confirms the complaint.
 
Event Description
Per clinical review: the manufacturer's clinical specialist spoke with the perfusionist regarding an incident the team had on (b)(4) 2020 with the yellow level sensor during a cardiopulmonary bypass (cpb) procedure.No other information was given at this point.The level sensing system during a procedure gave the team a level not attached message.It is unknown what reservoir they use, if they wait the five minutes prior to attaching the sensors to the level sensor adhesive pads, or if they had a concern about the gel that the sensors need attachment with.It is unknown about a delay or harm.There is routinely no blood loss due to this issue.There is typically no delay or harm due to this issue.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10924837
MDR Text Key218946623
Report Number1828100-2020-00464
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier0088799000717
UDI-Public(01)0088799000717(11)180202
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number195215
Device Catalogue Number195215
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2020
Initial Date FDA Received12/01/2020
Supplement Dates Manufacturer Received12/08/2020
12/08/2020
01/06/2021
02/15/2021
Supplement Dates FDA Received12/16/2020
12/18/2020
01/27/2021
02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-