• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W4TR01
Device Problems Electrical /Electronic Property Problem (1198); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 10/09/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 459888 lead, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy pacemaker's (crt-p) emergency ventricle-only pacing setting was inadvertently programmed on during a patient session resulting in extracardiac stimulation.The patient is to be brought in for reprogramming back to previous settings.The crt-p remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA QUAD CRT-P MRI SURESCAN
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10927023
MDR Text Key218914175
Report Number3004209178-2020-20860
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735644
UDI-Public00643169735644
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/28/2020
Device Model NumberW4TR01
Device Catalogue NumberW4TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2020
Date Device Manufactured05/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5086MRI58 LEAD, 5086MRI52 LEAD
Patient Outcome(s) Required Intervention;
Patient Age87 YR
-
-