VYAIRE MEDICAL 3100A HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100 A |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire field service went onsite, performed 7 year preventive maintenance based on manufacturing recommendation.The technician replaced the driver power module, all voltage readings and test were confirmed.The column filter was cleaned and the trap filters was replaced with non inventory parts.The unit was tested and verified operations okay.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the 3100 a unit shut off while on patient, staff heard alarm, unknown what alarms.The customer confirmed that there is no patient harm associated with this reported event.
|
|