Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the tip of the instrument was fractured.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned device found signs of repeated use and one of the spike tips has fractured.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|