MAUDE Adverse Event Report: ZIMMER BIOMET, INC. SPIKE ARM; CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. SPIKE ARM; CLAMP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the tip of the instrument was fractured.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned device found signs of repeated use and one of the spike tips has fractured.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIKE ARM
Type of Device
CLAMP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10932327
MDR Text Key219362549
Report Number0001822565-2020-03972
Device Sequence Number1
Product Code HXD
UDI-Device Identifier00889024227644
UDI-Public(01)00889024227644
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00599707300
Device Lot Number63596539
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A