MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL HEAD METAL ASR; METAL FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL HEAD METAL ASR; METAL FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD METAL ASR
Device Problem Naturally Worn (2988)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Asr legal notification received.It was reported that the patient had a series of check for the past years until x-ray and a check of cobalt-chromium shows elevated metal ion level and the implant was worn.The surgeon requested for revision however upon examination an aortic aneurysm was detected and it will not be possible to perform the revision.Doi: (b)(6) 2005 : dor: none reported (unk hip).
 
Manufacturer Narrative
Product complaint # (b)(4).No device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed andthe investigation will be re-opened as necessary.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : investigation summary: the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa-001226.Ongoing post market surveillance is conducted per our procedures for this product.Investigation section: previous investigations that have included device history reviews since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.Therefore, no dhr (device history record) review for this individual asr component will be carried out at this point in time.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.The investigation to determine root cause was conducted under mdd capa-001226.Ongoing post market surveillance is conducted per our procedures for this product.Root cause category: cause not established.E3 initial reporter occupation: lawyer.No 510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP FEMORAL HEAD METAL ASR
Type of Device
METAL FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10936675
MDR Text Key219308998
Report Number1818910-2020-25974
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD METAL ASR
Was Device Available for Evaluation? No
Date Manufacturer Received12/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
-
-