• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number 0375704500
Device Problems Insufficient Cooling (1130); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Burn(s) (1757); Burn, Thermal (2530)
Event Date 11/05/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the patient was burned.
 
Manufacturer Narrative
Alleged failure: shaver over heating burned salesforce case number (b)(4).Harm: burned pt.Delay: 5m.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) excessive force applied to the cutter/burr by the user with poor or no suction during use.(excessive force may cause friction due to bur shaft assembly and housing assembly interaction.) (2) tissue blocking the suction pathway would prevent hot fluid from being removed from the joint.(3) poor arthroscopy pump suction.(4) additionally, during arthroscopic procedures, it is common to use an rf energy wand.Inherent to use of an rf probe is the risk of potential superficial patient burns due to the heat generated from the rf output.Therefore use of the rf probe can also be considered a possible root cause as well.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
 
Event Description
It was reported that the patient was burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGING, FORMULA SHAVER (HAND CONTROL)
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10936986
MDR Text Key219297861
Report Number0002936485-2020-00503
Device Sequence Number1
Product Code GFA
UDI-Device Identifier07613327057638
UDI-Public07613327057638
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0375704500
Device Catalogue Number0375704500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Date Manufacturer Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-