SMITH & NEPHEW MEDICAL LTD. FOOTSWITCH, MULTI-FUNCTION, VERSAJET II; LAVAGE, JET Back to Search Results
Catalog Number 66800472
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during inspection the console was found with the foot pedal defective and needs replacement.No patient was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of functional failure could not be reproduced.Product passed functional testing with test console and pump cartridge.Probable cause maybe a connection issue.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTSWITCH, MULTI-FUNCTION, VERSAJET II
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10942248
MDR Text Key219485170
Report Number8043484-2020-04054
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
PMA/PMN Number
K143115
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800472
Device Lot Number05668
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-