Model Number 9-PDA-005
Device Problem Material Deformation (2976)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The investigation results will be provided in a subsequent submission.
 
Event Description
An 5mm amplatzer duct occluder was placed per ifu.The device appeared to "cobra head" and was removed.Device was replaced with a new 5mm amplatzer duct occluder.Patient was reported to be in stable conditon.
 
Manufacturer Narrative
An event of deformation upon deployment was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis; however, two images were received for analysis.Based solely on the aforementioned photos, an amplatzer duct occluder appeared to deploy with some degree of tilt out of the distal end of the sheath.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Type of Device
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10942770
MDR Text Key219512218
Report Number2135147-2020-00526
Device Sequence Number1
Product Code MAE
UDI-Device Identifier00811806011127
UDI-Public00811806011127
Combination Product (y/n)N
PMA/PMN Number
P020024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model Number9-PDA-005
Device Catalogue Number9-PDA-005
Device Lot Number7021548
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-