• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE AFFINITY FRAME; TABLE, OBSTETRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE AFFINITY FRAME; TABLE, OBSTETRIC Back to Search Results
Device Problem Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
The hillrom technical support representative performed troubleshooting over the phone with the account.The technician at the account isolated the left siderail and the bed stopped moving on its own.Per the hillrom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance (pm).Pm will minimize downtime due to excessive wear.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician at the account will replace the left siderail to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom technical support received a report from the account stating the bed hi low function was moving on its own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY FRAME
Type of Device
TABLE, OBSTETRIC
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key10949003
MDR Text Key220950946
Report Number1824206-2020-00497
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/25/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-