Results: the velocity was kinked approximately 8.0 cm from the hub.The device was ovalized approximately 155.0 cm and 159.0 cm from the hub.Conclusions: evaluation of the returned velocity revealed a proximal kink.If the velocity is forcefully manipulated at extreme angles during use, damage such as a kink may occur.Further evaluation of the device revealed distal ovalizations.This damage was likely incidental to the complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|