• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2020
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the key broke inside the key slot.Hospital staff and distributor staff attempted to retrieve the broken key from the slot, but were unsuccessful.
 
Manufacturer Narrative
Visual inspection of the driver confirmed the customer-reported issue of the key broken inside of the key switch.The driver history as well as documentation received from the customer indicated that the driver was functional upon arrival at customer site and damage did not occur during shipping.It cannot be conclusively determined what caused the key to break in the key switch.The potential for companion 2 keys to break is also a known issue that is being addressed in capa-0275, companion 2 broken keys.Syncardia has completed its investigation and is closing this file.Ce 5429 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10958614
MDR Text Key219943123
Report Number3003761017-2020-00254
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Date Manufacturer Received11/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-