• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.However, the customer stated that disconnecting the cooling gas solves the issue.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported to vyaire medical that the ventilator had constantly low pressure gas alarm and cannot be reset.At this time, there is no information regarding patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key10960490
MDR Text Key220245101
Report Number2021710-2020-12947
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003543
UDI-Public(01)10846446003543(11)20080909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/13/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/09/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-