ROCHE DIAGNOSTICS ELECSYS ANTI-SARS-COV-2; CORONAVIRUS SEROLOGICAL REAGENT Back to Search Results
Catalog Number 09203095190
Device Problems False Positive Result (1227); High Test Results (2457); Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The investigation determined the customer was not using the required rack adapters while using 13mm tubes and the customer did not perform level 2 qc.Product labeling states: "not using a roche rack cup adapter with 13 mm tubes or incorrect use of a roche rack cup adapter may cause incorrect results or errors.Always use roche rack cup adapters with 13mm tubes.".
 
Event Description
The initial reporter received questionable elecsys anti-sars-cov-2 results for 1 patient sample on a cobas e 411 immunoassay analyzer, serial number (b)(4).The initial result was coi 1.49 (positive).The repeated results were coi 0.192 and coi 0.732 (both negative).The results were not reported outside of the laboratory.
 
Manufacturer Narrative
The customer tested the sample a fourth time and the result was negative.A numerical value for this repeated result was not provided.The customer believed the correct result to be negative.Based on the provided data/information a general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-SARS-COV-2
Type of Device
CORONAVIRUS SEROLOGICAL REAGENT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10962162
MDR Text Key226993033
Report Number1823260-2020-03113
Device Sequence Number1
Product Code QKO
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Catalogue Number09203095190
Device Lot Number51233901
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-