• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER Back to Search Results
Model Number 33103
Device Problem Sharp Edges (4013)
Patient Problems Pain (1994); Skin Tears (2516); No Code Available (3191); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the patient experienced cuts in skin which caused pain and bleed because of the sharp plastic.
 
Manufacturer Narrative
The reported issue could not be confirmed as no sample was returned for evaluation.A potential failure mode for this could be "irregular cut" with potential root caus be "worn or damaged cutting knifes".It was unknown whether the device had met specifications.The product was used for treatment but it was unknown whether the product had caused the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "intended use: the self-adhering male external catheter is used for the drainage of urine.The catheter is applied by the patient or caregiver.Description / indication: the self-adhering male external catheter is designed for the management of adult male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Warning: reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury or illness of the patient.Precaution: do not use if allergic reaction occurs or if patient has known allergies to device components.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Follow directions to remove if experiencing swelling, numbness, discomfort, pain, discoloration, or abnormal appearance.Note: if experiencing problems with use of the device, please consult your healthcare professional for assistance.Directions to apply: 1) verify correct size prior to use.2) trim pubic hair if necessary.3) wash penis with mild soap and warm water.Dry thoroughly.Wear time may be reduced if skin is not dry or cream/oil is used.4) open package at perforation.Remove catheter from plastic insert, if present.5) place rolled end over the end of the penis, leaving a small space between the end of the penis and the cone of the catheter.6) unroll the catheter over penis.7) gently squeeze the catheter to properly seal adhesive to the skin.If possible, avoid leaving a rolled ¿collar¿ around the base of the penis.Important: wear time may be reduced if adhesive is not properly sealed to the skin.8) connect the catheter to a drainage system.Make sure to check that connections are secure before use.Directions to remove: 1) ensure drainage bag is empty.2) disconnect catheter from the drainage system.3) gently roll the catheter forward and off the penis.If necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.Disposal: after removal, dispose by placing the used catheter into a waste container." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient experienced cuts in skin which caused pain and bleed because of the sharp plastic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER
Type of Device
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10962692
MDR Text Key220188922
Report Number1018233-2020-21482
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741070686
UDI-Public(01)00801741070686
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number33103
Device Catalogue Number33103
Device Lot NumberJUDY0157
Was Device Available for Evaluation? No
Date Manufacturer Received02/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-