Model Number 71992-01
Device Problem High Readings (2459)
Patient Problems Hypoglycemia (1912); Cramp(s) (2193); Loss of consciousness (2418)
Event Date 11/21/2020
Event Type  Injury  
Manufacturer Narrative
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint, and there was no indication that the product did not meet specification.Dhrs (device history record) for the freestyle libre sensor and freestyle libre sensor kit were reviewed, and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported a high reading issue with the freestyle libre 2 sensor.The customer reported unspecified high readings, and symptoms of slow reaction, muscle cramps, and loss of consciousness.The customer was treated with glucagon and honey by a third-party, and an ambulance was called.A capillary reading of 4.1 mmol/l was obtained by paramedics, and the customer was given additional glucagon via iv.Comparison readings taken against built-in meter, when plotted on a parkes error grid, fell into the "c" zone, showing the difference in values to be clinically significant.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported a high reading issue with the freestyle libre 2 sensor.The customer reported unspecified high readings, and symptoms of slow reaction, muscle cramps, and loss of consciousness.The customer was treated with glucagon and honey by a third-party, and an ambulance was called.A capillary reading of 4.1 mmol/l was obtained by paramedics, and the customer was given additional glucagon via iv.Comparison readings taken against built-in meter, when plotted on a parkes error grid, fell into the "c" zone, showing the difference in values to be clinically significant.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10962850
MDR Text Key220121045
Report Number2954323-2020-12830
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2021
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2021
Date Manufacturer Received08/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight57
-
-