Model Number W1DR01
Device Problems Misconnection (1399); Device Contamination with Body Fluid (2317)
Patient Problems Arrhythmia (1721); Heart Block (4444)
Event Date 11/19/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported approximately two days post the implant of an right ventricular (rv) lead that the patient was experiencing arrhythmia and heart block.The lead exhibited no capture and on x-ray it was identified that the lead had dislodged.The lead was repositioned and when plugged into the device port up to four times there was no capture and a pacing issue was noted also.After multiple attempts the physician replaced the implantable pulse generator (ipg) and normal pacing was obtained.Fluid such as blood or tissue is suspected for the port issue.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURE XT DR MRI SURESCAN
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10967676
MDR Text Key220185080
Report Number3004209178-2020-21382
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier00643169634589
UDI-Public00643169634589
Combination Product (y/n)N
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/14/2022
Device Model NumberW1DR01
Device Catalogue NumberW1DR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Date Manufacturer Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5076-45 LEAD, 5076-52 LEAD.; 5076-45 LEAD, 5076-52 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age68 YR
-
-