Device was received and evaluated.Evaluation determined that the device ceramic insulation at the distal end of the inner sheath was observed to be broken.Approximately, based on the returned broken device, 30 percent of the insulation tip broken off from the device and observed with sharp edge.The reported failure mode had been addressed by the original equipment manufacturer.Solid mechanics was identified as a contributor to ceramic tip fracture.Device history records for this product were reviewed and showed the product met all specifications upon release.This unit was produced after corrective actions were taken, and had a new design tip with a black ceramic material.Based on the device evaluation results, tip fracture occurred to the device could result from excessive force or moment applied during use.As a ceramic material is brittle in nature, ceramic tips are susceptible to fracture even though a robust re-design with a more durable material has been made.The instructions for use (ifu) states "always keep sheaths parallel to one another when assembling and disassembling.If the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip.A broken tip, or fragments of a damaged tip, can potentially pass through the outer sheath and into the patient.If drag or resistance is encountered during assembly or disassembly, stop align working element and sheath parallel to one another before proceeding".Olympus will continue to monitor the field performance of this device.
|