The field service engineer (fse) conducted a site visit and was able to confirm the reported issue via visual inspection.The fse was unable to reproduce the reported issue due to a broken electrode wire.The issue was resolved by repairing the broken cable on the diluent electrode.The instrument was validated running a daily check and performing quality control (qc).The qc passed and was within published ranges.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review for similar complaints were performed for the serial number (b)(4) from 13oct2019 through aware date of (b)(6) 2020.There were no similar complaints identified during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: error message: [3000] diluent shortage.Cause: the level sensor of diluent is not detecting fluid.Measurement is suspended.Solution: measurement is automatically resumed after the diluent is replenished.If the diluent level is sufficient, contact tosoh service center or local representatives.The most probable cause of this reported event was the diluent electrode wire was broken.
|