Model Number 71364450
Device Problem Nonstandard Device (1420)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during a thr procedure, for a r3 straight shell impactor, the threads were bad and needed replaced.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The threading on the tip of the impactor is damaged rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 STRAIGHT SHELL IMPACTOR
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10980105
MDR Text Key220632547
Report Number1020279-2020-07319
Device Sequence Number1
Product Code LXH
UDI-Device Identifier03596010607034
UDI-Public03596010607034
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364450
Device Catalogue Number71364450
Device Lot Number16DSA0007
Was Device Available for Evaluation? No
Date Manufacturer Received12/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-