Model Number FOL0102
Device Problem Missing Information (4053)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the patient was confused by the packaging of foley 2 -way statlock whether if it was sterile or non- sterile which had no markings on it to determine sterility.Also informed caller this product was non-sterile.
 
Event Description
It was reported that the patient was confused by the packaging of foley 2 -way statlock whether if it was sterile or non- sterile which had no markings on it to determine sterility.Also informed caller this product was non-sterile.
 
Manufacturer Narrative
The reported event was inconclusive.It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.No sample was returned for evaluation.A potential root cause for this failure could be due to "lack of understanding of catheter".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A review of the device labeling is adequate to prevent the reported event."the statlock® device is a stabilization device for compatible catheters, and contains the following warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10993256
MDR Text Key220971672
Report Number1018233-2020-21617
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076114
UDI-Public(01)00801741076114
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2023
Device Model NumberFOL0102
Device Catalogue NumberFOL0102
Device Lot NumberJUEV0744
Was Device Available for Evaluation? No
Date Manufacturer Received05/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-