Catalog Number 397003-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion hospital cart will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The companion hospital cart was not supporting a patient.The customer, a syncardia certified hospital, reported that while clinical engineering team was conducting usual maintenance, staff noticed that the ground pin in the hospital cart was broken off.
|
|
Manufacturer Narrative
|
Physical inspection confirmed the male ground pin was attached to the female mating pin in the ac power cord.This damage can occur when the ac power cord is improperly disconnected from the hospital cart by applying force instead of releasing the cord using the red locking tab.The ground pin was easily removed from the cord receptacle by pulling back on the red release locking mechanism.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5436 follow-up report 1.
|
|