A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that post-aquablation procedure the patient was taken back to the operating room (or) because continuous bladder irrigation was prematurely stopped, and the patient began to bleed and developed a bladder tamponade (per manufacturer's instructions for use, bleeding is a potential perioperative risk of the aquablation procedure).The patient underwent clot evacuation and cauterization to address the event.No adverse health consequences were reported with the patient due to this event.No malfunction of the aquabeam robotic system was reported.
|
H3.Device evaluation by manufacturer: the aquabeam robotic system was not returned for investigation as it is currently in use at the user facility.A review of the aquabeam robotic system's log file was reviewed, which confirmed no malfunctions during the aquablation procedure.The review indicated that the system functioned as designed.A review of the device history record (dhr) for ab2000-b rev b/serial number (b)(6) was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the device met all design and manufacturing specifications when released for distribution.A review for similar complaints under ab2000-b/serial number (b)(6) confirmed six (6) other similar events.A review of similar complaints across all lots confirmed a total of 15 similar events.The aquabeam robotic system's instructions for use (ifu), ifu0104-00, rev.B, was reviewed and states the following: 4.3.Warnings: procedure.As with any surgical urologic procedure, potential perioperative risks of the aquablation procedure include: o bleeding.After aquabeam handpiece removal, follow the standard clot evacuation procedure to remove clots and tissue with a cystoscopic sheath by using an ellik bladder evacuator or toomey syringe.Use one of the following methods to achieve hemostasis: o non-resective bladder neck cautery followed by foley balloon catheter insertion.O under spinal anesthesia, insert a balloon catheter in the bladder with bladder neck traction, then fill the bladder with sterile saline and maintain for approximately 30-60 minutes before starting cbi (continuous bladder irrigation) balloon catheter in bladder with bladder neck traction.Balloon catheter in prostatic fossa: inflate balloon with 5cc in the bladder.Under trus guidance retract balloon into prostatic fossa.Inflate balloon to 30-50% of initial prostate volume.Apply mild traction on the catheter to hold the balloon catheter in place.Balloon catheter in bladder, no traction.Start cbi per hospital protocol.A root cause for the reported event could not be determined.The aquabeam robotic system's ifu lists bleeding as a potential risk of the aquablation procedure.Based on the information provided that continuous bladder irrigation (cbi) was prematurely discontinued post-aquablation procedure, plus the review of the log file, dhr and ifu, the event is considered not to be device related.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|