Model Number 2650
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2020
Event Type  malfunction  
Manufacturer Narrative
Aspen surgical received a report from the distributor indicating that product was found with seal issues.The actual device was returned for evaluation.Photographic evidence along with the manufacturing lot number were also provided for review.The distributor indicated that the defects were found during incoming inspection.A review of the photos and sample confirmed the issue from the distributor.Analysis of the finished good lot number was reviewed.No non-conformance's were noted during the manufacturing process.According to the manufacturing process, these parts are loaded into recessed pockets on a packaging machine per specified work instructions.Patient labels and rulers are loaded into the machine by an automated auto-loader, which dispenses each component into the pocket.If incorrectly loaded or not orientated properly, the product interferes with the sealing process.The product that was evaluated was determined that the machine caused the parts to drop from too high of a height.This caused the product to not be placed correctly.Therefore, the likely root cause is attributed to machine issue.This issue was improved with an update to the pneumatic cylinder to allow for more control of the drop height into the pocket for better product placement.Based on this information, no further action is required.
 
Event Description
Aspen surgical received a report from the distributor indicating that a surgical marking pen was found with defective seals.The item was not in use.No injury/death was reported.Customer reported multiple lots for the same issue.The lot numbers and their respective complaint numbers are as follows: (b)(4), lot 233293.(b)(4), lot 223954.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RICHARD-ALLAN REGULAR TIP PEN W/ RULER AND 6-UP LABELS
Type of Device
MARKING PEN
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
Manufacturer (Section G)
SAME AS ABOVE
Manufacturer Contact
jordan hackert
6945 southbelt dr. s.e.
caledonia, MI 49316
6165367508
MDR Report Key11002210
MDR Text Key223623145
Report Number1836161-2020-00061
Device Sequence Number1
Product Code FZZ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2650
Device Lot NumberSEE ABOVE
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-