The hillrom technician found the siderail interface board, motor control board and left siderail cable needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in july 2020.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the siderail interface board, motor control board and left siderail cable to resolve the issue.Based on this information, no further action is required.
|