ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Model Number 03-2722-9 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances, or abnormalities identified during the manufacturing process, which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed, and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached, and a cause could not be confirmed.
|
|
Event Description
|
A user facility clinic manager (cm) reported that a blood leak occurred with a combiset at the beginning of a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed by a patient care technician (pct).When a member of the nursing staff attempted to inspect the heparin line and address the leak, the heparin line separated from the combiset and the blood leak increased.There was no defect or damage noted on the combiset.It was not reported by the user facility that the patient experienced a serious injury, adverse event, or required medical intervention as a result of this blood leak.The patient¿s estimated blood loss (ebl) was approximately 300 ml.The patient was restarted on a new machine, and treatment completed successfully with new supplies.The complaint sample was discarded, and is not available to be returned to the manufacturer for physical evaluation.
|
|
Search Alerts/Recalls
|
|
|