Model Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances, or abnormalities identified during the manufacturing process, which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed, and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached, and a cause could not be confirmed.
 
Event Description
A user facility clinic manager (cm) reported that a blood leak occurred with a combiset at the beginning of a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed by a patient care technician (pct).When a member of the nursing staff attempted to inspect the heparin line and address the leak, the heparin line separated from the combiset and the blood leak increased.There was no defect or damage noted on the combiset.It was not reported by the user facility that the patient experienced a serious injury, adverse event, or required medical intervention as a result of this blood leak.The patient¿s estimated blood loss (ebl) was approximately 300 ml.The patient was restarted on a new machine, and treatment completed successfully with new supplies.The complaint sample was discarded, and is not available to be returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11014654
MDR Text Key221762490
Report Number8030665-2020-01838
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model Number03-2722-9
Device Catalogue Number03-2722-9
Device Lot Number19KR01101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received11/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS DIALYZER
Patient Age72 YR
Patient Weight84
-
-