C.R. BARD, INC. (BASD) -3006260740 HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, 14.5F
|
Back to Search Results |
|
Model Number 5833730 |
Device Problems
Obstruction of Flow (2423); Improper Flow or Infusion (2954); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Thrombosis/Thrombus (4440)
|
Event Date 08/02/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 02/2021) the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that three months post catheter placement in the right internal jugular vein, the catheter allegedly had functional problems and poor blood drainage.It was further reported that there was clotting in the tube.Reportedly, the patient has intermitted bleeding problem.Current status of patient is unknown.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a complaint history review was performed.This is the fifth complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the device was not returned for evaluation.The investigation is inconclusive for the reported poor function, poor blood drainage, poor dialysis effect, and easy clotting of the tubing issue.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2021).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that three months post catheter placement in the right internal jugular vein, the catheter allegedly had functional problems and poor blood drainage.It was further reported that there was clotting in the tube.Reportedly, the patient has intermittent bleeding problem.Current status of patient is unknown.
|
|
Search Alerts/Recalls
|
|
|