Catalog Number 8065752450
Device Problems Excess Flow or Over-Infusion (1311); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that irrigation solution of infusion did not stop because of closing failure of the clamp during surgery.The surgery was completed after replacing the product with another one.There's no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The used infusion manifold connected to the infusion cannula line was visually inspected and no obvious defects were detected.The sample was tested with the lab stock components using a constellation console.The sample could prime, tune with the ultrasonic handpiece and pass intraocular pressure calibration successfully.The snap clamp could be closed to stop fluid and opened to allow fluid flowing in several times without any issue.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11019922
MDR Text Key222643705
Report Number1644019-2020-00690
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number8065752450
Device Lot Number2388071H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Date Manufacturer Received01/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-