A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing printouts and reproduced error by attempting a sample run.Fse inspected sample nozzle on main arm and found serum buildup inside nozzle tip.Fse resolved the complaint by cleaning the sample nozzle and instructed customer on how to perform cleaning.Fse validated instrument by successfully completing quality control run without error and results were within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The aia-2000 operator's manual under appendix 4: error messages states the following: a64 main arm blocked: error detected in suction operation of loader (and beltline) specimen.The following message will also be output simultaneously.[2070] clogging detected during specimen suction by main arm.Cause: the negative pressure detected after specimen suction exceeded the standard.The specified.Amount of specimen may not have been obtained because the sampling nozzle was blocked.The measurement result will be flagged (sc flag).Solution: verify that the specimen is free of solid substances (such as fibrin) or that there is sufficient volume of specimen if it is prepared in the primary tube and retry the measurement.If retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event was due to clogged sample nozzle.
|
A customer reported getting sampling error "a64 main arm blocked" and "2070 clogging detected during specimen suction by main arm" error message on the aia-2000 instrument.Technical support specialist (tss) instructed the customer to check the waste chute and customer confirmed it was three-fourths full.Instrument is down.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), prolactin (prl) and follicle stimulating hormone (fsh).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|