Model Number V60
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.(b)(6) 2020.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Event Description
It was reported to philips that the device came from another depot and that the battery is not charging.The device was not in clinical use at the time of the event.There was no report of patient or user harm.The device was evaluated by the customer with assistance from a philips remote service engineer (rse).The customer connected to ac and verified it was not charging.The customer stated the device would not power on at this point on ac or dc.The customer verified that the battery was from 2019.The battery charging voltage is.007 connected to ac battery.Once it was disconnected the unit was off.The rse advised the customer to verify the voltage coming from the power supply and found no voltage.The customer stated 110 vac at the power cord going into the unit.And did not have meter leads to verify 110 vac to the input of the power supply.The customer was going to verify if the power supply is getting the 110 vac input and call back.
 
Manufacturer Narrative
G4:08jan2021.B4:13jan2021.Multiple good faith efforts were made to obtain information regarding predefined battery questions, device evaluation, repair, and operational status with no response from the reporter and therefore cannot be determined.If additional information is later obtained, the complaint will be reopened and a supplemental report will be submitted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11022816
MDR Text Key222317150
Report Number2031642-2020-04595
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/19/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-