SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the freedom driver did not maintain the proper normotensive pressure values on the patient simulator.
 
Manufacturer Narrative
High left arterial pressure (lap) can be indicative of a faulty piston and cylinder assembly (pca).During investigation testing, a known functioning pca was installed and the driver was retested.With the new pca installed, the driver passed all functional test steps.The root cause of the reported issue that the freedom driver did not maintain the proper normotensive pressure values, high left arterial pressure (lap), was determined to be a malfunction of the piston cylinder assembly.Syncardia has a corrective and preventative action (capa) for the issue of freedom driver elevated left arterial pressure (lap) during testing.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key11028537
MDR Text Key222082701
Report Number3003761017-2020-00262
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-