Model Number 595000-001 |
Device Problems
Noise, Audible (3273); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the freedom driver exhibited an unusual noise while supporting the patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently changed to a backup driver.
|
|
Manufacturer Narrative
|
The complaint was confirmed and replicated during testing when a grinding sound was heard coming from the driver.The root cause of the 'funny' noise reported by the customer is binding motors within the primary motor gear box assembly.Syncardia has a corrective and preventive action (capa) for the issue of freedom driver primary motor and piston cylinder assembly.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|