Model Number 1457Q/75
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problems Bradycardia (1751); Weakness (2145)
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that a patient presented in the emergency room with weakness and bradycardia.External pacer pads were applied.Interrogation showed that the patient's right ventricular and left ventricular leads had dislodged.The leads were explanted and replaced.Related manufacturer reference number: 2017865-2020-23041.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 75 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11028861
MDR Text Key222081585
Report Number2017865-2020-23047
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734510172
UDI-Public05414734510172
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model Number1457Q/75
Device Catalogue Number1457Q/75
Device Lot NumberA000066139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient Weight85
-
-