V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020, date of report: 17dec2020.
 
Event Description
The customer reported blower overheating alarm when powered on.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.The customer advised there were no significant errors in the logs.Philip's remote service engineer (rse) advised the caller to perform a full performance verification test (pvt) to help diagnose the issue.
 
Manufacturer Narrative
Following the evaluation and testing of the device, the customer was unable to duplicate the error.The device passed testing.No other abnormality was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11032908
MDR Text Key222691801
Report Number2031642-2020-04623
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-