Model Number V60 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020, date of report: 17dec2020.
|
|
Event Description
|
The customer reported blower overheating alarm when powered on.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.The customer advised there were no significant errors in the logs.Philip's remote service engineer (rse) advised the caller to perform a full performance verification test (pvt) to help diagnose the issue.
|
|
Manufacturer Narrative
|
Following the evaluation and testing of the device, the customer was unable to duplicate the error.The device passed testing.No other abnormality was observed.
|
|