WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD CUTTER; CUTTER, WIRE Back to Search Results
Model Number 388.72
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2020
Event Type  malfunction  
Manufacturer Narrative
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020, the patient underwent an unknown procedure.During the procedure, the rod cutter broke.There was no fragment generated from the broken device.The procedure was successfully completed with no surgical delay.The patient's status was okay.This report involves 1 rod cutter.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part number: 388.72.Lot number: t135084.Manufacturing site: tuttlingen.Release to warehouse date: 29-mar-2016.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The raw material certificate was reviewed and the used material was according to the specification of the device.Visual inspection: the rod cutter (p/n: 388.72, lot #: t135084) was returned and received at us customer quality (cq).Upon visual inspection, it was observed that the distal tip was broken and fragments are not returned.No other issues were observed with the returned device.Device failure/defect identified? yes.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Furthermore, the complaint relevant dimensions cannot be checked for dimensional accuracy, because of the design of the device.Service and repair evaluation: the customer reported that the rod cutter broke.The repair technician reported that the blades of cutter are damaged and pieces are missing.Damaged component is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Document/specification review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed - rod cutter.Complaint confirmed? yes, the device received was broken.Hence confirming the allegation.Investigation conclusion: the complaint condition was confirmed for the rod cutter (p/n: 388.72, lot #: t135084).There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD CUTTER
Type of Device
CUTTER, WIRE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11036266
MDR Text Key222313808
Report Number2939274-2020-05643
Device Sequence Number1
Product Code HXZ
UDI-Device Identifier10705034773759
UDI-Public(01)10705034773759
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number388.72
Device Catalogue Number388.72
Device Lot NumberT135084
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Date Manufacturer Received02/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-