• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problems Signal Artifact/Noise (1036); Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the left ventricular unipolar vector showed poor sensing and baseline noise when evaluating waveform on freeze capture but all other parameters were normal.The patient was brought in for additional isometric testing but no issues were found and parameters were normal.No programing changes were made except changes to ensure the electrocardiogram line graphs could be more visible.The patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11041471
MDR Text Key222621725
Report Number2017865-2020-23135
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2013
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot Number3148410
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/17/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
-
-