SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL Back to Search Results
Model Number 71441144
Device Problems Crack (1135); Nonstandard Device (1420)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during a tka procedure, using a gii mis dcf align gde, the tab on top was loose and didn't lock the distal cut block on , the surgeon had to hold distal guide on by hand, the procedure was completed successfully with no significant delay.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.The clinical/medical team concluded, per complaint details, the both bumpers broke during femoral impaction.Reportedly, all pieces were retrieved by hand and the procedure was completed within a 0-30 minute surgical extension using a backup device without patient injury or other complications.Patient impact beyond the 0-30 minute surgical delay and the modified surgical procedure with use of a backup device would not be anticipated as the procedure was reportedly completed without patient injury.No further medical assessment is warranted at this time.A visual inspection confirmed the journey fem impact bumper lt is cracked.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF ALIGN GDE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11041966
MDR Text Key222613471
Report Number1020279-2020-07590
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497161
UDI-Public03596010497161
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441144
Device Catalogue Number71441144
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received02/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-