Visual inspection of the device showed dried saline on the distal end and attached to the inner wall of the irrigation tubing inside the luer fitting.Dried body fluid found on the handle, main body tubing and on the distal end.Electrical continuity checks revealed no electrical opens or shorts, as checked manually using a multi-meter and breakout box.All electrodes, sensor and thermocouple resistances measured in spec and were typical.During functional testing the steering knob and the tension control knob functioned properly on both lock and unlock positions.The device's lumen was leak tested, pressure decay values were 0.1048 psi, 0.0822 psi, and 0.0668 psi.These values are within an acceptable limit.The device was connected to a metriq pump to purge saline for approximately 30 seconds and verify all ports were functioning.The distal tip electrode was immersed in 0.45% saline for 30 minutes.Tc+ and tc- to tip resistance measurements remained open.Next, the device was connected to a metriq pump.After irrigating for approximately 30 minutes at a flow rate of 30 ml/min, tc+ and tc- to tip resistance measurements remained open.A metriq pump was connected to the device and the distal end was suspended in the center of a 250ml beaker during the purge function and saline appear to be coming out of all ports as expected.Then changed metriq pump flow rate to 30ml/min (ablation rate), and after 5 minutes the empty beaker contained approximately 150 ml of saline, which was within specifications.The device was connected to a maestro generator, rhythmia hdx system.Impedance readings were normal.Three 120 second, 50w ablations were performed.All three ablations (straight, right & left curve) were normal with no error codes or warnings.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|