Model Number 802100 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) failed.As a result, an alternate device was employed.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
Updated blocks: h3 and h6.The reported complaint was not verifiable.Multiple diligence attempts for part return were unsuccessful so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Per the manufacturer's sales associate, when the customer turned the pump on, everything came on except for the touch screen.The pump was turned off and back on, but had the same issue.The monitor cables were disconnected and reconnected with no resolution.There were no audible tones and the screen remained a dark grey color with no activity.After examination, the monitor had a frayed cable and that may have been the cause of the screen failure.
|
|
Event Description
|
Additional information received that the surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|