ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Failure to Cut (2587)
Patient Problem Abrasion (1689)
Event Date 11/12/2020
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.(b)(6).
 
Event Description
It was reported that during surgery the device is not cutting properly.There was a five minute delay, and an additional skin graft was needed.No additional consequences have been reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d8, d9, g3, g6, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined the rpms were out of specification on the low end, calibration was out at the zero reading and the width plates were rough.The motor, width plates, gears, spring seal, vespel and semi-bearings and other parts were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key11058221
MDR Text Key223426922
Report Number0001526350-2020-01098
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number30729200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
-
-