Catalog Number A1114 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that they experienced difficulty opening the mayfield infinity skull clamp back up during an unspecified procedure.The surgeon stated it was locked very well, and they had to loosen the screw on the side to release it.No patient injury, or surgical delay was reported.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).The mayfield infinity skull was returned for evaluation: the complaint is confirmed via inspection.Unit received worn, requiring replacement of worn internal parts, general maintenance, and cleaning.This unit is beyond integra's 7 years recommended life cycle (manufactured in 2006).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|