Model Number 9-ASD-032
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Embolism (1829)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
The reported event of device migration could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2020, a 32 mm amplatzer septal occluder was selected for implant.During the implant procedure the minnesota maneuver was successfully used to implant the device.Upon releasing the device from the delivery cable, the device migrated into the left atrium.The device was successfully snared and retrieved percutaneously to resolve the event.The patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key11061251
MDR Text Key223451056
Report Number2135147-2020-00551
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010236
UDI-Public00811806010236
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number7411845
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-