MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
A follow up will be submitted when additional information becomes available.
 
Event Description
It was reported that the cardiohelp showed the alarm ¿pump disposable error¿ during patient treatment.No indication of actual or potential for harm or death was reported.Complaint number: (b)(4).
 
Manufacturer Narrative
During patient transfer a "pump disposable error " message was reported.The device caused the complaint and was not able to work within factory specifications.As stated by the service and sales unit on 2020-12-16, the device had no malfunction.No service will be performed as the problem has been solved by phone.With reference to the current risk analyses u7.2.4.1 and u7.2.5.1 the following most probable root causes could be determined: user does not perceive or recognize how to correctly mount the transport guard the transport guard are mounted incorrectly or not mounted at all no coupling or insufficient coupling between the cardiohelp device and the disposable during transport.Additional condition: the mechanical impact on the cardiohelp device and disposable is too high.The ifu(instructions for use | 1.8 | en | 09) of the cardiohelp includes a warning under chapter 2 "do not remove the disposable product during normal operation." furthermore in the ifu (instructions for use | 1.8 | en | 09) it is stated in chapter 2.1.4 "please refor tho the respective instruction for use of the disposables".In reference of the current ifu for disposables (instructions for use | 1.5 | g-270 | 02) the following is stated in chapter 5.3.1 safety instructions for the oxygenator: incorrect installation of the hls module advanced can lead to device malfunction.This can endanger the patient.Use the device only together with the device cardiohelp-i.Install or remove the device only when the pump of the cardiohelp-i is at a standstill (0 rpm).Ensure that the device is fitted onto the device correctly and securely fixed, to eliminate the risk of magnetic decoupling between the drive and the centrifugal pump.Based on this the reported failure "pump disposable error" could be confirmed, but not a product related malfunction.The device history record (dhr) of the cardiohelp (material: 70104.8012, serial: (b)(6) for which a customer complaint was received, was reviewed on 2021-01-18.The dhr does not show any abnormality or issue that is related or can have led to the customer complaint.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required the occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key11063602
MDR Text Key225385011
Report Number8010762-2020-00455
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP
Device Catalogue Number70104.8012
Date Manufacturer Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-