ST. JUDE MEDICAL FLEXABILITY ABLATION CATHETER, BI-D, CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number A701125
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/05/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned due to a temperature issue.The temperature accuracy read as 192.1°c and was not within specification.Microscopic analysis revealed char on the distal tip.Further investigation revealed it is possible the high temperature is related to incorrect length of the pi tubing; however, it could not be confirmed due to the adhesive around the pi tubing.
 
Event Description
This report is to advise of an event observed during analysis confirming a high ablation temperature noted while testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXABILITY ABLATION CATHETER, BI-D, CURVE D-F
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11064511
MDR Text Key223404950
Report Number3008452825-2020-00719
Device Sequence Number1
Product Code OAD
UDI-Device Identifier05415067011114
UDI-Public05415067011114
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P110016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model NumberA701125
Device Catalogue NumberA701125
Device Lot Number6822866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient Weight90
-
-