Model Number 801763
Device Problems Display or Visual Feedback Problem (1184); Failure to Power Up (1476)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
Per the subsidiary, the roller pumps were able to be started but it was not possible to program the pumps to obtain the desired flows for the patient without the ccm booting up.
 
Event Description
It was reported that the central control monitor (ccm) would not turn on.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was not verifiable.Multiple diligence attempts for part return and additional information were unsuccessful so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Additional information was received that the reported issue occurred during priming of the device for a cardiopulmonary bypass (cpb) procedure.The surgical procedure was completed successfully.There was a five minute delay to change out the batteries.There was no blood loss nor adverse consequences to the patient.
 
Manufacturer Narrative
Per the subsidiary, when the heart lung machine (hlm) was disconnected from alternating current (a/c) power it shut down.The user facility installed new batteries that were from another manufacturer.After reconnecting the hlm, the unit was able to be used for the surgical procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11065768
MDR Text Key225728789
Report Number1828100-2020-00500
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K151349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-