• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 106531US
Device Problems Display or Visual Feedback Problem (1184); Output Problem (3005)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2020
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
It was reported that the patient's controller had a broken display screen.They were unable to view pump numbers or see any display on the controller.The patient was thenon the power module and the screen and numbers were within normal limits.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of difficulty in reading device parameters on lcd screen was confirmed.The log file spanned approximately 16 days ((b)(6) 2020 to (b)(6) 2020 per the timestamp).There were no events related to the log file.Initial evaluation of the returned hm3 system controller, serial (b)(6), revealed a damaged lcd screen that made device parameters difficult to read.The lcd was replaced with a test lcd resolving the issue.The controller was functionally tested and was connected to a mock circulatory loop for an extended period of time without any issue.The damaged lcd did not affect the functionality of the controller.A root cause of the reported event was not determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or qa specifications.(b)(6) was shipped to the customer on 31jul2018.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Heartmate iii instructions for use section 8 entitled ¿equipment storage and care¿ and heartmate iii patient handbook section 6 entitled ¿caring for the equipment¿ addresses how to properly care for, maintain, and store the equipment for proper use.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 SYSTEM CONTROLLER
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
MDR Report Key11067542
MDR Text Key223728875
Report Number2916596-2020-06080
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013235
UDI-Public00813024013235
Combination Product (y/n)N
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/24/2021
Device Model Number106531US
Device Catalogue Number106531US
Device Lot Number6468192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age24 YR
Patient Weight101
-
-