• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Model Number 03-2742-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event Description
A user facility clinic manager reported a blood leak that occurred at the end of the patient¿s hemodialysis (hd) treatment.The blood leak was visually observed at the connection where the bloodline connects to the transducer.There were no loose connections observed.There was no defect or damage noticed on the combiset.The patient¿s estimated blood loss (ebl) was approximately 300 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient completed treatment without changing machines or supplies.The complaint device was not returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11074175
MDR Text Key223789676
Report Number8030665-2020-01892
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100309
UDI-Public00840861100309
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number03-2742-9
Device Catalogue Number03-2742-9
Device Lot Number20LR01240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received12/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient Weight94
-
-